Celltac α+ MEK-1303 from Nihon Kohden is an automatic analyser that integrates haematology (with 3-part leukocyte differentiation), CRP determination and HbA1c testing in a single compact instrument. The system maximises testing efficiency through a single aspiration process, providing combined haematology results including CRP from a single sample.
In certain situations, a CRP test can be more useful than an ESR test, because the markers used in each test behave differently depending on the stage and cause of inflammation. MEK-1303 provides complete haematology results, including CRP, through a single aspiration process.
Monitoring HbA1c is essential for diabetic patients. MEK-1303 provides individual HbA1c results (in HbA1c measurement mode only the HbA1c result is displayed) and is NGSP certified.
This unique technology reduces the “re-entry” of blood cells after passing through the detection aperture: the swirling flow pushes the cells towards the drainage path. The effect is particularly useful for samples with a low cell volume, ensuring accurate CBC results.
| Reportable parameters (22) | WBC, LY%, MO%, GR%, LY#, MO#, GR#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PCT, MPV, PDW, P-LCR, CRP, HbA1c |
|---|---|
| Research parameters (4) | P-LCC, Mentzer Index, RDWI, NLR |
| Measurement modes | Open and Closed |
| Measurement time | CBC: <1 min (open) / <1 min 30 s (closed) · CRP: <3 min 40 s · HbA1c: <5 min 30 s |
| Sample volume (normal mode) | CBC: 20 µl · CBC + CRP: 26 µl · HbA1c: 10 µl |
| Measurement methods | WBC, RBC, PLT — detection by electrical resistance; HGB — colorimetric; HCT — calculated from the RBC histogram; WBC differentiation — calculated from the WBC histogram |
| Parameter | Reproducibility | Linearity |
|---|---|---|
| WBC | ≤2.0% | ±3.0% or ±0.30 ×10³/µl |
| RBC | ≤1.5% | ±3.0% or ±0.08 ×10⁶/µl |
| HGB | ≤1.5% | ±1.5% or ±0.2 g/dl |
| HCT | ≤1.5% | ±3.0% or ±1.0% |
| MCV / MCH / MCHC | ≤1.0% / ≤2.0% / ≤2.0% | — |
| PLT | ≤4.0% | ±10.0% or ±20 ×10³/µl |
| CRP | — | ±15% or ±0.1 mg/dl |
| HbA1c | — | ±10.0% |
| Dimensions (W × D × H) | 230 × 450 × 428 mm |
|---|---|
| Weight | 22 kg |
| Voltage | AC 100–240 V |
| Frequency | 50 or 60 Hz |
| Power consumption | 150 VA |
| External outputs | LAN × 1, USB × 2, RS-232C × 3 |