Celltac α+ MEK-1303 from Nihon Kohden is an automatic analyser that integrates haematology (with 3-part leukocyte differentiation), CRP determination and HbA1c testing in a single compact instrument. The system maximises testing efficiency through a single aspiration process, providing combined haematology results including CRP from a single sample.

Haematology with CRP

In certain situations, a CRP test can be more useful than an ESR test, because the markers used in each test behave differently depending on the stage and cause of inflammation. MEK-1303 provides complete haematology results, including CRP, through a single aspiration process.

  • CRP as a rapid marker of inflammation.
  • Closed Mode for safe operation.

Individual HbA1c test

Monitoring HbA1c is essential for diabetic patients. MEK-1303 provides individual HbA1c results (in HbA1c measurement mode only the HbA1c result is displayed) and is NGSP certified.

DynaHelix Flow technology

This unique technology reduces the “re-entry” of blood cells after passing through the detection aperture: the swirling flow pushes the cells towards the drainage path. The effect is particularly useful for samples with a low cell volume, ensuring accurate CBC results.

Other features (MEK-1300 series)

  • Barcode reader for QC and reagent management.
  • Touch panel for fast operation.
  • Connection to a PCL printer.
  • Smart ColoRerun Assist for easy alarm recognition.
  • ASTM connection between the LIS and the MEK-1300 series.

Consumables and calibrators

  • CR-420W — 50 CRP reagent cartridges.
  • HA-420W — 50 HbA1c reagent cartridges.
  • CR-CAL — CRP calibrator (0.25 mL × 3 vials).
  • HA-410V CAL — HbA1c calibrator (0.25 mL × 3 vials).
  • MEK-CAL — haematology calibrator.
  • MEK-3CL / 3CN / 3CH — quality control for CBC and CRP.
  • WA-130W — external printer (optional).

Key specifications

Reportable parameters (22)WBC, LY%, MO%, GR%, LY#, MO#, GR#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PCT, MPV, PDW, P-LCR, CRP, HbA1c
Research parameters (4)P-LCC, Mentzer Index, RDWI, NLR
Measurement modesOpen and Closed
Measurement timeCBC: <1 min (open) / <1 min 30 s (closed) · CRP: <3 min 40 s · HbA1c: <5 min 30 s
Sample volume (normal mode)CBC: 20 µl · CBC + CRP: 26 µl · HbA1c: 10 µl
Measurement methodsWBC, RBC, PLT — detection by electrical resistance; HGB — colorimetric; HCT — calculated from the RBC histogram; WBC differentiation — calculated from the WBC histogram

Reproducibility and linearity

ParameterReproducibilityLinearity
WBC≤2.0%±3.0% or ±0.30 ×10³/µl
RBC≤1.5%±3.0% or ±0.08 ×10⁶/µl
HGB≤1.5%±1.5% or ±0.2 g/dl
HCT≤1.5%±3.0% or ±1.0%
MCV / MCH / MCHC≤1.0% / ≤2.0% / ≤2.0%
PLT≤4.0%±10.0% or ±20 ×10³/µl
CRP±15% or ±0.1 mg/dl
HbA1c±10.0%

Physical specifications

Dimensions (W × D × H)230 × 450 × 428 mm
Weight22 kg
VoltageAC 100–240 V
Frequency50 or 60 Hz
Power consumption150 VA
External outputsLAN × 1, USB × 2, RS-232C × 3